Electrode placement, Electrode placement -3, Table 5-1 – Physio-Control LIFEPAK 20e User Manual

Page 101

Advertising
background image

5

P

ad
dle Acc

essor

y

Options

Paddle Accessory Options

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

5-3

©2006-2013 Physio-Control, Inc.

FAST-PATCH electrodes can be used on pediatric patients if the placement meets conditions
noted in the following paragraphs. Usually, these conditions can be met by patients weighing
15 kg (33 lb) or more.

Electrode Placement

When using QUIK-COMBO or FAST-PATCH electrodes, ensure that the electrodes:

• Fit completely on the chest as described on

page 3-4

or

page 4-3

.

• Have at least 2.5 cm (1 in.) of space between electrodes.

• Do not overlap bony prominences of sternum or spine.

To place the electrodes:

1 Prepare the patient for electrode placement:

• Remove all clothing from the patient’s chest.
• Remove excessive chest hair as much as possible. Avoid nicking or cutting the skin if using

a shaver or razor. If possible, avoid placing the electrodes over broken skin.

• Clean and dry the skin. If there is ointment on the patient’s chest, use soap and water to

clean the skin. Briskly wipe the skin dry with a towel or gauze. This mildly abrades the skin
and removes oils, dirt, and other debris for better electrode adhesion to the skin. Do not use
alcohol, tincture of benzoin, or antiperspirant to prepare the skin.

2 Slowly peel back the protective liner on the electrodes, beginning with the cable connection

end (refer to

Figure 5-2

).

Figure 5-2

Peeling the Liner from the Electrode

3 Place the electrodes in the anterior-lateral or anterior-posterior position, as described on

page 3-4

or

page 4-3

, depending on the therapy to be provided and special placement

considerations.

4 Starting from one edge, firmly press the electrode on the patient’s chest to eliminate air

pockets between the gel surface and the skin. Firmly press all adhesive edges to the skin.

Note: Once applied, therapy electrodes should not be repositioned.

Table 5-1

QUIK-COMBO Electrodes

Type

Description

QUIK-COMBO

Electrodes, with .6 m (2 ft) of lead wire, designed for
patients weighing 15 kg (33 lb) or more.

QUIK-COMBO - RTS

Electrodes, providing a radio-transparent electrode and
lead wire set, designed for patients weighing 15 kg (33 lb)
or more.

QUIK-COMBO with REDI-PAK™
preconnect system

Electrodes designed for patients weighing 15 kg (33 lb) or
more and allow preconnection of the electrode set to the
device while maintaining electrode shelf life and integrity.

Pediatric QUIK-COMBO

Electrodes designed for patients weighing 15 kg (33 lb) or
less.