Cable connection, Ecg monitoring and therapy procedures – Physio-Control LIFEPAK 20e User Manual

Page 102

Advertising
background image

Paddle Accessory Options

5-4

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

Cable Connection

To connect QUIK-COMBO electrodes to the QUIK-COMBO therapy cable:

1 Open the protective cover on the QUIK-COMBO therapy cable connector (refer to

Figure 5-3

).

2 Insert the QUIK-COMBO electrode connector into the therapy cable connector by aligning the

arrows and pressing the connectors firmly together for proper attachment.

Figure 5-3

Connecting QUIK-COMBO Electrodes to Therapy Cable

To properly connect FAST-PATCH electrodes to the FAST-PATCH defibrillation adapter cable
and ensure energy delivery:

1 Attach the defibrillation cable to the electrode post (before applying electrodes to the patient,

when possible).

2 Support the electrode post when attaching the defibrillation cable (refer to

Figure 5-4

). Firmly

press the cable onto the electrode until a click is heard or felt.

Figure 5-4

Connecting FAST-PATCH Electrodes to Defibrillation Cable

3 Pull up gently on the connector to confirm that the defibrillation cable is securely connected to

the electrode.

Note: If you are reattaching the defibrillation cable to an electrode that is already on the
patient, lift the adhesive edge under the electrode post slightly and place your finger under the
post. Connect the cable as described in the preceding steps.

ECG Monitoring and Therapy Procedures

For adult patients, follow the procedures for ECG monitoring, AED defibrillation, manual
defibrillation, synchronized cardioversion, and pacing described in

Section 3

or

Section 4

.

WARNINGS!

Possible patient skin burns.

Do not use Pediatric QUIK-COMBO electrodes on adults or larger children. Delivery of
defibrillation energies equal to or greater than 100 joules (typically used on adults) through these
smaller electrodes increases the possibility of skin burns.

Possible pediatric patient skin burns.

Noninvasive pacing may cause patient skin irritation and burns, especially with higher pacing
current levels. Inspect underlying skin of the

♥ electrode frequently after 30 minutes of

continuous pacing. Discontinue noninvasive pacing if skin burn develops and another method of
pacing is available. On cessation of pacing, immediately remove or replace electrodes with new
ones.

Therapy cable
connector

QUIK-COMBO

electrode connector

Advertising