Appendix A

Specifications and Performance Characteristics

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

A-1

©2006-2013 Physio-Control, Inc.

All specifications are at 20°C (68°F) unless otherwise stated.

GENERAL

The LIFEPAK 20e defibrillator/monitor has seven main operating modes:

Manual Mode

Provides normal operating capability for ALS users

AED Mode

Provides normal operating capability for BLS users

Archive Mode

Allows operator to print, edit or delete previous patient
records

Setup Mode

Allows operator to configure the instrument

Service Mode

Allows operator to execute device diagnostic tests and
calibrations

Inservice Mode

Provides simulated waveforms for demonstration
purposes

Auto Test
Mode

Provides daily automatic tests of critical circuits

POWER

AC Powered

100-120 VAC 50/60 HZ, 220-240 VAC 50/60 HZ. Total
power draw less than 120 Volt-Amperes (VA), or less
than 150 VA with the CodeManagement Module
installed.

Operating Time

A new, fully charged, internal backup battery provides a minimum of the following prior to
shutdown:

---total--

--after low battery--

Monitoring plus SpO

2

(minutes)

210

5

Defibrillation (360 J
discharges):

140

3

Monitoring Plus Pacing
(minutes at 100 mA, 60 ppm)
Plus SpO

2

:

110

2

Battery Charging Time

Battery charges while defibrillator operates from AC
power. Typical battery charge time of less than four
hours when defibrillator is powered off and AC power is
applied.

Low Battery Indication

Low battery message in status area and warning tone
sounds.

Battery Status Indicator

Indicates available battery capacity for defibrillator.

A Service Indicator

When Error Detected.