ALARMS SETUP MENU

The Alarms

Setup menu allows you to define alarms and set the alarm volume level. When you

select a menu item, the screen displays a help message. The underlined options are factory
default settings.

PRINTER SETUP MENU

The Printer

Setup menu allows you to define automatic event printing and ECG frequency

response. When you select a menu item, the screen displays a help message. The underlined
options are factory default settings.

Table 8-15

Alarms Setup Menu

Menu Item

Help Message

Options

VOLUME

Set volume for alarms, tones,
and voice prompts

Select volume level from gradient display.
The minimum setting reduces but does not
silence alarms.

ALARMS

Enable alarms at power up

ON

: Enables alarms for Heart Rate, SpO

2

,

and CO

2

whenever defibrillator power is

turned on.

OFF

: Alarms available through

ALARMS

button.

VF/VT ALARM

Alarms when VF or VT
detected

ON

: Enables VF/VT alarm whenever

defibrillator power is turned on.

OFF

: VF/VT alarm available through

ALARMS

button.

Table 8-16

Printer Setup Menu

Menu Item

Help Message

Options

AUTO PRINT

Specify Auto Print Event

Refer to

Table 8-17

.

ECG MODE

Default ECG frequency
response

MONITOR

or

DIAGNOSTIC

MONITOR MODE

Default monitor frequency
response for printer and
display

1–30 HZ

or

.5–40 HZ

.

DIAGNOSTIC MODE

Default diagnostic frequency
response for printer

.05–40 HZ

or

.05–150 HZ

.

ALARM EVENTS

Print waveforms with alarm
events and print with
CODE SUMMARY report

ON

or

OFF

.

EVENT
WAVEFORMS

Print waveforms with user-
entered events and print with
CODE SUMMARY report

ON

or

OFF

.