Co2 detection, Cleaning, Cleaning -21 – Physio-Control LIFEPAK 20e User Manual

Page 67: Detection

Advertising
background image

3

M

o

n

itoring

Monitoring

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

3-21

©2006-2013 Physio-Control, Inc.

CO

2

Detection

A CO

2

waveform appears when any CO

2

is detected, but CO

2

must be greater than 3.5 mmHg for

a numerical value to be displayed. However, the CO

2

module will not recognize a breath until the

CO

2

is at least 8 mmHg (1.0% or 1 kPa). Valid breaths must be detected in order for the apnea

alarm to function and to count the respiratory rate (RR).

When CO

2

is not detected in the cardiac arrest situation—for example, the CO

2

waveform is

either dashes “---” or a flat solid line at or near zero—several factors must be quickly evaluated.
Assess for the following causes:

Equipment Issues

• FilterLine set is disconnected from the endotracheal tube (ETT)

• System is purging due to fluid in the patient/sensor connection from ET administration of

medications

• System is auto-zeroing

• FilterLine connection to device is loose

Loss of Airway Function

• Improper placement of ETT

• ETT dislodgment

• ETT obstruction

Physiological Factors

• Apnea

• Loss of perfusion

• Exsanguination

• Massive pulmonary embolism

• Inadequate CPR

Cleaning

Accessories for CO

2

monitoring are disposable and are intended for single-patient use. Do not

clean and reuse a FilterLine set. Dispose of the contaminated waste according to local protocols.

Advertising