Screen Messages

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

C-1

©2006-2013 Physio-Control, Inc.

Table C-1

Summary of Screen Messages

Message

Description

ABNORMAL ENERGY
DELIVERY

A discharge occurred when the paddles were shorted together (refer
to warning,

page 4-15

); an open air discharge occurred; or, the patient

impedance is out of range. This message may also appear in certain
types of internal faults.

ACCESS DENIED

Three consecutive incorrect passcode attempts were entered.

AED MODE

The device is monitoring the patient condition and functioning as a
semiautomatic external defibrillator.

ALARMS SILENCED

The alarms are silenced. An alert tone and this message appear
periodically to remind you that alarms have been silenced.

ANALYZING NOW–
STAND CLEAR

The AED is analyzing the patient ECG rhythm.

C LEAD OFF

The ECG electrode “C” is disconnected.

CHARGING TO XXX J

The front panel or the standard paddles

CHARGE

button was pressed.

CHECK CO2

EtCO

2

FilterLine tubing or exhaust port is blocked.

CHECK FOR PULSE

The AED prompt that appears after each standard 3-shock sequence
or

NO SHOCK ADVISED

message.

CHECK PATIENT!

A potentially shockable rhythm is detected when the VF/VT alarm is
on.

CHECK PRINTER

The printer door is open; there is no paper in the printer; or, there is
another printer malfunction.

CO2 AUTOZERO

EtCO

2

monitor is automatically performing a zero-point calibration.

CO2 FILTERLINE
BLOCKAGE

EtCO

2

FilterLine tubing is kinked or clogged; the message appears

after 30 seconds of unsuccessful purging.

CO2 FILTERLINE OFF

EtCO

2

FilterLine tubing is disconnected or is not securely connected to

the device.

CO2 FILTERLINE
PURGING

EtCO

2

FilterLine tubing is kinked or clogged with liquid.

CO2 INITIALIZING

EtCO

2

monitor is performing a self-check.

CONNECT CABLE

In manual mode, the therapy cable was not connected when you
pressed

CHARGE

. In pacing mode, the QUIK-COMBO cable was not

connected when you increased current. In AED advisory mode, the
QUIK-COMBO cable was not connected when you pressed

ANALYZE

.

CONNECT
ELECTRODES

The defibrillator detects that the therapy electrodes are disconnected.