Lifepak, 20e defibrillator, 20e defibrillator/monitor operator’s checklist – Physio-Control LIFEPAK 20e User Manual

Page 187

Advertising
background image

LIFEPAK

®

20e Defibrillator/Monitor Operator’s Checklist

Unit Serial No.:________________________

Location: ____________________________

Instruction

Recommended

Corrective Action

Date

Initials

Insert a

in the box after completing each instruction.

1 Check printed result of 3 A.M. daily auto test. (If no self-test results have

printed, refer to Step 10.) If:

SELF-TEST FAILED

Contact qualified service personnel.

SELF-TEST DID NOT COMPLETE –
PERFORM MANUAL TEST

Connect test plug or secure paddles in well,
and perform manual User Test. If

CONNECT

TO TEST PLUG

message appears, contact

qualified service personnel.

2 Inspect physical condition for:

Foreign substances

Clean the defibrillator.

Damage or cracks

Contact qualified service personnel.

3 Inspect power source for:

AC power connector plugged into unit and
AC power source; AC Mains LEDs on
defibrillator and optional
CodeManagement Module are lit

If AC Mains LEDs are not lit, contact qualified
service personnel.

Broken, loose, or worn power cable

Replace damaged or broken parts.

4 Check therapy and ECG electrodes for:

Use By date

Replace if past expiration date.

Spare electrodes available

Obtain spare electrodes.

5 Examine accessory cables for:

Cracking, damage, broken or bent parts
or pins, and paddle surfaces for pitting

Replace damaged or broken parts.

6 Disconnect the defibrillator from AC power, wait 2 seconds, press ON and check for:

Momentary

SELF-TEST

messages,

illumination of LEDs, and speaker beep

If absent, contact qualified service personnel.

Service LED is lit

If lit, power off device, then power back on.
If service LED stays lit, contact qualified
service personnel.

7 Check ECG printer for:

Adequate paper supply

Replace if necessary.

Ability to print

If not working, contact qualified service
personnel.

8 Confirm therapy cable connected to defibrillator and perform cable check:

If QUIK-COMBO

®

therapy cable is connected:

• Confirm test plug connected to therapy cable.

• Press

ANALYZE

button.

• After

ANALYZING NOW

message, look

for

REMOVE TEST PLUG

message.

If

CONNECT CABLE

or

CONNECT

ELECTRODES

message appears, contact

qualified service personnel.

If HARD PADDLES are connected:

• Confirm paddles properly seated in wells.

• Select 10 J and press the

CHARGE

button.
Note: Discharging > 10 joules in paddle
wells may damage defibrillator

If

CONNECT CABLE

or

CONNECT

ELECTRODES

message appears, contact

qualified service personnel.

• When fully charged, press the

SHOCK

button on the paddles and look for the

ENERGY DELIVERED

message.

If no energy delivered and shock tone
continues, press Speed Dial to disarm, and
contact qualified service personnel.

9 Reconnect to AC power and then power off the device.

Daily inspection and testing of the defibrillator/monitor using
this Operator’s Checklist is recommended. This form may
be reproduced.