Side view of codemanagement module, Side view of codemanagement module -15, Warning – Physio-Control LIFEPAK 20e User Manual

Page 37

Advertising
background image

2

B

asic Orien

tation

Basic Orientation

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

2-15

©2006-2013 Physio-Control, Inc.

System Connector

For LIFEPAK 20e devices that do not have a CodeManagement Module attached, the system
connector is used to transfer setup information to another LIFEPAK 20e device or connect to
LIFENET

®

Device Agent. For LIFEPAK 20e devices that have a CodeManagement Module

attached, the system connector is used to connect to LIFENET Device Agent.

ECG/Sync Connector

The ECG/Sync connector provides remote synchronization and real-time ECG output to a third
party monitor.

CO

2

Exhaust Port

The CO

2

exhaust port connects to a scavenger system when monitoring EtCO

2

during use of

anesthetics.

Side View of CodeManagement Module

The CodeManagement Module has the following indicators and connectors on the right side.

Figure 2-18

Side View of CodeManagement Module

The following paragraphs provide additional information about the side view.

TrueCPR Device Port

The TrueCPR device port is used to transfer data from the TrueCPR™ Coaching Device to the
LIFENET System. Only the TrueCPR device should be connected to this port. Refer to

"Data

Transfer from TrueCPR Device," page 6-13

.

Note: The TrueCPR device may not be available in all countries. Contact your local
Physio-Control representative for more information.

WARNING!

Shock hazard.

All equipment connected to the system or ECG/sync connector must be battery powered or
electrically isolated from AC power according to EN 60601-1. For more information, contact
Physio-Control Technical Support.

TrueCPR

device port

Wireless

indicators

Wireless

On/Off switch

On

Off