LIFEPAK 20e Defibrillator/Monitor Operating Instructions

4-19

©2006-2013 Physio-Control, Inc.

4 Press

SYNC

. Confirm the sync LED blinks with each detected QRS complex.

Note: Press

SYNC

again to deactivate synchronous mode.

5 Observe the ECG rhythm. Confirm that a triangle sense marker appears near the middle of

each QRS complex. If the sense markers do not appear or are displayed in the wrong
locations (for example, on the T-wave), select another lead. (It is normal for the sense marker
location to vary slightly on each QRS complex.)

6 Prepare the patient’s skin for therapy electrode application as described in Step 3 of

Defibrillation Procedure

,

page 4-16

.

7 Connect the therapy electrodes to the therapy cable, and confirm cable connection to the

defibrillator.

8 Apply therapy electrodes to the patient in the anterior-lateral or anterior-posterior position. If

using standard paddles, apply conductive gel to the paddles and place paddles on the
patient’s chest.

9 Press

ENERGY SELECT

.

10 Press

CHARGE

.

Note: If the

REMOVE TEST PLUG

message appears, disconnect test plug and connect

therapy electrodes to QUIK-COMBO therapy cable and press

CHARGE

.

11 Make certain all personnel, including operator, stand clear of the patient, bed, and any

equipment connected to the patient.

12 Confirm ECG rhythm. Confirm available energy.

13 Press and hold the

SHOCK

button on the defibrillator until the

ENERGY DELIVERED

message

appears on the screen. For standard paddles, press and hold both

SHOCK

buttons on the

paddles simultaneously until the

ENERGY DELIVERED

message appears on the screen.

Release buttons. For safety reasons, the

SHOCK

button on the defibrillator is disabled when

using standard paddles.

Note: To disarm (cancel a charge), press the Speed Dial. The defibrillator disarms

automatically if shock buttons are not pressed within 60 seconds, or if you change the
energy selection after charging begins. Press

CHARGE

to restart charging.

14 Observe patient and ECG rhythm. Repeat procedure from Step 4, if necessary.

WARNING!

Possible lethal arrhythmia.

Ventricular fibrillation may be induced with improper synchronization. DO NOT use the ECG
from another monitor (slaving) to synchronize the LIFEPAK 20e defibrillator/monitor discharge.
Always monitor the patient’s ECG directly through the ECG cable, therapy cable or use the
remote synchronization procedure. Confirm proper placement of the sense markers on the ECG.