Troubleshooting tips for data transmission, Troubleshooting tips for data transmission -14 – Physio-Control LIFEPAK 20e User Manual

Page 122

Advertising
background image

Data Management

6-14

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

TROUBLESHOOTING TIPS FOR DATA TRANSMISSION

Table 6-4

Troubleshooting Tips for Data Transmission

Observation

Possible Cause

Corrective Action

1

SEND DATA

option does

not appear in Options/
Archives screen.

CodeManagement Module not
properly connected to
defibrillator.

• Contact qualified service

personnel.

2

TRANSMISSION FAILED

message appears.

No wireless connection to
LIFENET System.

• Verify Wireless switch on

side of CodeManagement
Module is in the

ON

position.

See

"Side View of

CodeManagement Module,"
page 2-15

.

• Contact your facility’s

LIFENET System
administrator to verify the
receiving computer is
operational.