Shock Advisory System

E-2

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

were selected to be a representative sample of rhythms seen during treatment of cardiac arrest.
The SAS Test Set contains the following ECG samples:

• 168 each coarse ventricular fibrillation (VF) (>= 200 µV peak-to-peak amplitude)

• 29 each fine ventricular fibrillation (< 200 and >= 80 µV peak-to-peak amplitude)

• 65 each shockable ventricular tachycardia (VT) (HR > 120 bpm, QRS duration >= 160 ms, no

apparent P waves, patient reported to be pulseless by the paramedics)

• 43 each asystole (< 80 µV peak-to-peak amplitude)

• 144 each normal sinus rhythm (NSR) (sinus rhythm, heart rate 60-100 bpm)

• 531 each other organized rhythm (includes all rhythms except those in other listed categories)

• 2 each transitional (transition occurs within the sample from nonshockable to nonshockable or

vice versa)

• 5 each shockable rhythms with pacemaker artifact (the pacemaker artifact is spread over time

by the filtering in the LIFEPAK 5 defibrillator)

• 2 each nonshockable rhythms with pacemaker artifact (the pacemaker artifact is spread over

time by the filtering in the LIFEPAK 5 defibrillator)

1

Each sample is run 10 times asynchronously.

2

Association for the Advancement of Medical Instrumentation. DF80: 2003 Medical electrical equipment-Part 2-4.
Arlington, VA: AAMI, 2004

3

Automatic External Defibrillators for Public Access Defibrillation: Recommendations for Specifying and Reporting
Arrhythmia Analysis Algorithm Performance, Incorporating New Waveforms, and Enhancing Safety. American Heart Associa-
tion (AHA) Task Force on Automatic External Defibrillation, Subcommittee on AED Safety and Efficacy. Circulation. 1997: Vol.
95: 1677-1682.

VF = ventricular fibrillation; VT = ventricular tachycardia; NSR = normal sinus rhythm

Table E-1

LIFEPAK 20e Series Overall SAS Performance

SAS Overall Performance

Sensitivity

>90%

Specificity

>95%

Positive Predictive Value

>90%

False Positive Rate

<5%

Table E-2

LIFEPAK 20e Series SAS Performance by Rhythm Category

Rhythm Class

ECG Test

1

Sample Size

Performance

Goal

Observed Performance

Shockable: Coarse VF

168

>90% sensitivity

LIFEPAK 20e meets the AAMI

2

DF80

requirements, EN60601-2-4:2003
requirements, and AHA

3

recommendations

Shockable: VT

65

>75% sensitivity

LIFEPAK 20e meets the AAMI DF80
requirements, EN60601-2-4:2003
requirements, and AHA recommendations

Nonshockable: NSR

144

>99% specificity

LIFEPAK 20e meets the AHA
recommendations

Nonshockable: asystole

43

>95% specificity

LIFEPAK 20e meets the AAMI DF80
requirements, EN60601-2-4:2003
requirements, and AHA recommendations

Nonshockable: all other
rhythms

531

>95% specificity

LIFEPAK 20e meets the AAMI DF80
requirements, EN60601-2-4:2003
requirements, and AHA recommendations

Intermediate: fine VF

29

Report only

>75% sensitivity