Physio-Control LIFEPAK 20e User Manual

Page 155

Advertising
background image

Appendix A

Specifications and Performance Characteristics

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

A-3

©2006-2013 Physio-Control, Inc.

Lead Selection

Leads I, II, III (3-wire ECG cable)

Leads I, II, III, AVR, AVL, AVF, and C acquired
simultaneously (5-wire ECG cable)

ECG Size

4, 3, 2.5, 2, 1.5, 1, 0.5, 0.25 cm/mV

Heart Rate Display

20–300 bpm digital display

Out of range indication: Display symbol "---"

Heart symbol flashes for each QRS detection

Continuous Patient
Surveillance System
(CPSS)

In AED mode, while Shock Advisory System is not active,
CPSS monitors the patient, via QUIK-COMBO paddles or
Lead II ECG, for potentially shockable rhythms.

Voice Prompts

Used for selected warnings and alarms (configurable on/off).

Analog ECG Output

1 V/mV x 1.0 gain <35 msec delay

Common Mode
Rejection

90 dB at 50/60 Hz

SpO

2

Masimo

®

sensors

Saturation range

1 to 100%

Saturation Accuracy

70–100% (0–69% unspecified)

Adults/Pediatrics

±2 digits (during no motion conditions)
±3 digits (during motion conditions)

Neonates

±3 digits (during no motion conditions)
±3 digits (during motion conditions)

Dynamic signal
strength bar graph

Pulse tone at the onset
of the pleth waveform

SpO

2

Update

Averaging Rate

User selectable 4, 8, 12 or 16 seconds

SpO

2

measurement

Functional SpO

2

values are displayed and stored

Pulse rate range

25 to 240 pulses per minute

Pulse rate accuracy

Adults/Pediatrics/
Neonates

±3 digits (during no motion conditions)
±5 digits (during motion conditions)

SpO

2

waveform with

autogain control

Alarms

Quick Set

Activates alarms for all parameters

VF/VT Alarm

Activates continuous CPSS monitoring in Manual Mode