Monitoring

3-24

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

15 CO

2

does not appear on

screen when FilterLine is
connected.

CodeManagement Module not
properly connected to
defibrillator.

• Contact qualified service

personnel.

Low battery voltage.

• Connect to AC power.

Defective battery in
CodeManagement Module.

• Contact qualified service

personnel.

CodeManagement Module
malfunction.

• Contact qualified service

personnel.

Table 3-4

Troubleshooting Tips for EtCO

2

(Continued)

Observation

Possible Cause

Corrective Action