Monitoring with the patient ecg cable, Monitoring with the patient ecg cable -5 – Physio-Control LIFEPAK 20e User Manual

Page 51

Advertising
background image

3

M

o

n

itoring

Monitoring

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

3-5

©2006-2013 Physio-Control, Inc.

3 Apply the therapy electrodes or standard paddles in the anterior-lateral position.

For therapy electrodes, confirm that the package is sealed and the Use By date has not

passed. For standard paddles, apply conductive gel over the entire electrode surface.

4 Connect the disposable therapy electrodes to the therapy cable.

5 Select

PADDLES

lead.

Monitoring with the Patient ECG Cable

There are two ECG cables available for ECG monitoring as shown in

Figure 3-2

: the 3-wire and

5-wire cables.

Connecting the Patient ECG Cable

Connect the cable by inserting the main cable connector into the green electrically isolated ECG
connector on the monitor.

Figure 3-2

3-wire and 5-wire ECG Cables

ECG Monitoring Procedure

1 Press

ON

.

2 Attach the ECG cable to the monitor.

3 Identify the appropriate electrode sites on the patient as shown in

Figure 3-3

.

Figure 3-3

Electrode Placement for ECG monitoring

4 Prepare the patient’s skin for electrode application:

• Shave excessive hair at electrode site. Avoid locating electrodes over tendons and major

muscle masses.

• For oily skin, clean skin with an alcohol pad.
• Dry the site with a brisk rub.

5-Wire cable

3-Wire cable

AHA Labels

IEC Labels

RA

Right Arm

R

Right

LA

Left Arm

L

Left

RL*

Right Leg

N

Negative

LL

Left Leg

F

Foot

C*

Chest

C

Chest

*Note: Not used for 3-wire cable.

LA/L

RA/R

LL/F

RL/N

C/C

Advertising