Setup Options

8-8

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

CPR METRONOME SETUP MENU

The CPR Metronome Setup menu allows you to enable the CPR metronome for AED mode, and
define the C:V (compression to ventilation) ratio for each Age-Airway setting. The C:V ratio is
adjustable to provide different settings based on patient age and airway status and according to
local medical protocols.

PRESHOCK
CPR

Set CPR interval after shock
advised decisions

OFF, 15, 30 SECONDS

Table 8-7

AED Mode Energy Protocol Setup Menu

Menu Item

Help Message

Options

PRESET
PROTOCOLS

Select a preset energy
protocol

Energy 1:

150

,

175

,

200

,

225

,

250

,

275

,

300

,

325

,

360

Energy 2:

150

,

175

,

200

,

225

,

250

,

275

,

300

,

325

,

360

Energy 3:

150

,

175

,

200

,

225

,

250

,

275

,

300

,

325

,

360

Energy 2 cannot be less than Energy 1.
Energy 3 cannot be less than Energy 2.

FLEXIBLE
PROTOCOLS

Repeat previous energy after

NO SHOCK ADVISED

ON

or

OFF.

STACKED
SHOCKS

Enable consecutive shocks
without CPR

ON

or

OFF.

Table 8-8

CPR Metronome Setup Menu

Menu Item

Help Message

Options

METRONOME

Enable metronome during
CPR

ON

or

OFF

(Turns metronome on or off for AED mode.)

ADULT – NO
AIRWAY

Set Adult – No Airway
compression/ventilation ratio

30:2

,

16:1

,

15:2

,

12:1

,

10:1

,

100:0

ADULT –
AIRWAY

Set Adult – Airway
compression/ventilation ratio

30:2

,

16:1

,

15:2

,

12:1

,

10:1

,

100:0

YOUTH – NO
AIRWAY

Set Youth – No Airway
compression/ventilation ratio

30:2

,

16:1

,

15:2

,

12:1

,

10:1

,

100:0

YOUTH –
AIRWAY

Set Youth – Airway
compression/ventilation ratio

30:2

,

16:1

,

15:2

,

12:1

,

10:1

,

100:0

Table 8-6

AED Mode CPR Setup Mode (Continued)

Menu Item

Help Message

Options

Advertising