Physio-Control LIFEPAK 20e User Manual

Page 98

Advertising
background image

Therapy

4-28

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

5 Monitor screen displays

distortion while pacing.

ECG electrodes not optimally
placed with respect to pacing
electrodes.

• Reposition electrodes away

from pacing electrodes.

Patient response to pacing is
highly variable with respect to
capture threshold and ECG
distortion.

• Select another lead (I, II, or

III).

• Consider changing pacing

rate.

6 Capture does not occur

with pacing stimulus.

Current (mA) set too low.

• Increase pacing current.

(Administer sedation/
analgesia as needed.)

7

CONNECT ELECTRODES

message appears.

Pacing cable or electrode
disconnected.

• Reconnect and set current.

Electrodes not adhering to skin. • Prepare skin.

Electrodes outdated.

• Apply new electrodes and set

current.

8 Pacing stops

spontaneously and

PACER

FAULT

message appears.

Internal error detected.

• Cycle power and start pacing

again.

• Obtain service by qualified

service personnel.

9

REMOVE TEST PLUG

message appears.

Test plug connected to therapy
cable.

• Disconnect test plug from

therapy cable.

• Connect electrodes to

therapy cable.

10 Intrinsic QRS complexes

not sensed when pacing.

ECG amplitude too low to be
sensed.

• Select another lead.

Intrinsic QRS complexes are
occurring during pacemaker's
refractory period.

• Adjust pacing rate.

11 Pacing starts

spontaneously.

Patient’s heart rate falls below
set pacing rate.

• Appropriate pacemaker

function assess patient.

During standby pacing, ECG
lead disconnects.

• Reconnect ECG lead.

12 Set pacing rate (ppm)

and ECG paced rate do
not appear to match.

Internal error detected.

• Print ECG and calculate the

pace rate.

13 Improper sensing

(for example, sensing on
T-waves).

QRS complex too small.
T-wave too large.

• Select another lead.

Table 4-4

Troubleshooting Tips for Noninvasive Pacing (Continued)

Observation

Possible Cause

Corrective Action

Advertising