Managing archived patient records, Entering archives mode, Printing archived patient reports – Physio-Control LIFEPAK 20e User Manual

Page 115

Advertising
background image

6

D

ata Mana

g

e

ment

Data Management

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

6-7

©2006-2013 Physio-Control, Inc.

MANAGING ARCHIVED PATIENT RECORDS

When you turn off the LIFEPAK 20e defibrillator/monitor, the current Patient Record is saved in
the archives. The following options are available for managing archived Patient Records:

• Print archived patient reports

• Transmit archived patient records

• Edit archived patient records

• Delete archived patient records

To perform any or all of these options, you must first enter the archives mode and then proceed
with the desired option.

ENTERING ARCHIVES MODE

Note: When you enter the archives mode, patient monitoring ends (for example, no ECG, no
alarms) and the current Patient Record is saved and closed.

PRINTING ARCHIVED PATIENT REPORTS

To enter the archives mode:
1 Press

OPTIONS.

2 Select

ARCHIVES.

3 Select

YES

to enter the patient

archives.

YES

closes and saves the current

Patient Record and ends patient
monitoring.
–or–
SELECT

NO

to clear the overlay

and return to the previous screen.

To print:
1 Be sure that you are in the archives

mode (refer to

"Entering Archives

Mode," page 6-7

).

2 Select

PRINT

.