Setup Options

8-6

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

To activate, select

ENERGY PROTOCOL

in the Paddle Default menu. Auto energy sequences

are disabled if you press the

ENERGY SELECT

control or change to or from AED mode during

use.

Table 8-4

Manual Mode Energy Protocol Setup Menu

Menu Item

Help Message

Options

PRESET
PROTOCOL

Select preset energy
protocol

FULL RANGE

,

PEDIATRIC

.

ENERGY 1

Select energy level for
shock 1

Full range:

100, 125, 150, 175

,

200

,

225,

250, 275, 300, 325, 360.

Pediatric:

2, 3, 4

,

5

,

6, 7, 8, 9, 10, 15, 20, 30,

50, 70, 100

.

ENERGY 2*

Select energy level for
shock 2

Full range:

100, 125, 150, 175, 200, 225,

250, 275

,

300

,

325, 360.

Pediatric:

2

,

3

,

4

,

5

,

6

,

7

,

8

,

9

,

10

,

15

,

20

,

30

,

50

,

70

,

100

.

ENERGY 3*

Select energy level for
shock 3

Full range:

100, 125, 150, 175, 200, 225,

250, 275, 300, 325

,

360

.

Pediatric:

2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30,

50, 70, 100, 125, 150.

*ENERGY 2 cannot be less than ENERGY 1

.

ENERGY 3 cannot be less than ENERGY 2.