Specifications and Performance Characteristics

A-6

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

Waveform Shape and Measured Parameters

Notes:
T1 = duration of Phase 1 in milliseconds
T2 = duration of Phase 2 in milliseconds

Patient

Impedance (

)

I1 (A)

I2 (A)

I3 (A)

I4 (A)

T1 (ms)

T2 (ms)

25

70.9

28.0

-28.0

-15.0

5.5

3.6

50

38.7

19.7

-19.7

-12.6

7.2

4.8

75

26.6

15.5

-15.5

-10.8

8.5

5.7

100

20.3

12.8

-12.8

-9.4

9.5

6.3

125

16.4

10.9

-10.9

-8.4

10.3

6.8

150

13.7

9.6

-9.6

-7.5

10.9

7.3

175

11.8

8.5

-8.5

-6.9

11.5

7.7

Note: Table values are nominal for a 360-joule shock.

T1

T2

I1

I2

I3

I4