Spo2 monitoring procedure, Spo2 waveform, Spo2 volume – Physio-Control LIFEPAK 20e User Manual

Page 58: Monitoring procedure, Waveform, Volume

Advertising
background image

Monitoring

3-12

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

SpO

2

Monitoring Procedure

The defibrillator controls power to the pulse oximeter. When the defibrillator is turned on, the
oximeter turns on and performs a self-test that requires up to 10 seconds. When the defibrillator
is turned off, the oximeter also turns off.

To conserve battery power, the pulse oximeter goes into “sleep mode” when not in use. Sleep
mode is activated within 10 seconds of disconnecting the sensor. During sleep mode, the screen
does not display SpO

2

information. The oximeter returns to normal mode after detecting a sensor

or a patient signal. The oximeter performs the self-test when it returns from sleep mode to active
mode.

The pulse oximeter measures SpO

2

levels between 1% and 100%. When SpO

2

levels are

between 70% and 100%, oximeter measurements are accurate within ±3 digits. When the pulse
oximeter measures SpO

2

levels less than 50%, the display shows < 50%.

To measure the patient’s SpO

2

levels:

1 Connect

the

SpO

2

cable to the monitor.

2 Attach the sensor to the SpO

2

cable and the patient.

3 Press

ON.

4 Observe the pulse bar for fluctuation. Amplitude of the pulse bar indicates relative signal

quality.

5 Adjust sensitivity, averaging time, and SpO

2

volume as necessary.

SpO

2

Waveform

The SpO

2

waveform can be displayed on waveform Channel 2 by selecting waveform Channel 2

and then selecting SpO

2

from the Waveform menu. The SpO

2

waveform automatically sizes itself

to provide optimum waveform viewing.

SpO

2

Volume

To adjust the pulse tone volume, highlight and select

SPO2

on the home screen.

The following overlay appears:

1 Highlight

and

select

SPO2 VOLUME

.

2 Rotate the Speed Dial to the desired

volume.

3 Press the Speed Dial to set the

volume.