8

S

etu

p

Opti

ons

Setup Options

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

8-5

©2006-2013 Physio-Control, Inc.

MANUAL MODE SETUP MENU

The Manual Mode Setup menu allows you to define defibrillation and synchronized cardioversion
settings. When you select a menu item, the screen displays a help message. The underlined
options are factory default settings.

Table 8-2

Manual Mode Setup Menu

Menu Item

Help Message

Options

SYNC

Set up sync defaults

Refer to

Table 8-3

.

PADDLES
DEFAULT

Default energy for paddles
or QUIK-COMBO
electrodes

Power-on energy setting for standard
paddles and therapy electrodes:

2, 5, 10,

50, 100, 125, 150, 175, 200

(joules), or

ENERGY PROTOCOL

.

ENERGY
PROTOCOL...

Energies for energy
protocol

Refer to

Table 8-4

.

INTERNAL
DEFAULT

Default energy for internal
paddles

Power-on energy setting for internal
paddles:

2

,

5, 10, 20, 30,

or

50

.

VOICE PROMPTS

Voice prompts active in
manual mode

ON

: Voice prompts active.

OFF

: Voice prompts inactive.

SHOCK TONE

Tone when defibrillator is
fully charged

ON

: A tone sounds.

OFF:

No tone.

MANUAL ACCESS

Means for accessing
manual mode

MANUAL / DIRECT:

No restrictions to

manual mode.

AED / DIRECT:

No restrictions to AED

mode.

AED / CONFIRM

:

Confirmation required to

gain manual access.

AED / PASSCODE

:

Passcode required to

enter manual mode.

SET PASSCODE...

Passcode required to
enter manual mode

If configured for passcode access:
None: Default passcode enabled.
New: User-defined 4-digit code enabled.

Table 8-3

Synchronization Defaults

Menu Item

Help Message

Options

SYNC AFTER
SHOCK

Resume sync after energy
transfer

ON

: Defibrillator returns to synchronous

mode after energy transfer.

OFF

: Defibrillator returns to asynchronous

mode.

REMOTE SYNC

Allow sync with remote
monitor

ON

: Remote synchronization active with

connection to a compatible remote monitor.

OFF

: No remote monitor connection active.

Advertising