LIFEPAK 20e Defibrillator/Monitor Operating Instructions

6-13

©2006-2013 Physio-Control, Inc.

Figure 6-3

IrDA Connections

DATA TRANSFER FROM TrueCPR DEVICE

Note: The TrueCPR device may not be available in all countries. Contact your local
Physio-Control representative for more information.

You can transfer data from the Physio-Control TrueCPR device to Physio-Control post-event
review tools using the TrueCPR device port on the CodeManagement Module. To transfer data:

1 Confirm the LIFEPAK 20e defibrillator/monitor is either turned off or in archives mode.

2 Connect an approved USB cable between the TrueCPR device and the CodeManagement

Module.

Note: Use only a USB cable that meets the following specifications: USB 2.0 A-Male to
Mini-B 5-pin Male, 28/24 AWB cable with ferrite core (gold plated), 1 m (3 ft).

3 Turn on the TrueCPR device and follow the instructions in the TrueCPR device Instructions

for Use.

Notes:

• It may take up to one minute for the TrueCPR device to display the Data Transfer screen.

• Data transfer takes approximately 3–5 minutes.

• If you attempt to transmit data from the LIFEPAK 20e defibrillator/monitor and the TrueCPR

device simultaneously, the data from the defibrillator will be transmitted first.

IrDA

Adapter/

Computer

Maximum Distance: 1.0 m (3.28 feet)

Defibrillator