Therapy

4-24

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

8

ABNORMAL ENERGY
DELIVERY

message appears

and Shock XJ Abnormal
annotated on printout.

Open air discharge with
standard paddles.

• Press paddles firmly on

patient’s chest when
discharging.

• Perform test discharges with

defibrillation checker.

Discharge occurs with
standard paddles shorted
together.

• Refer to warning,

page 4-15

.

Patient impedance out of
range.

• Increase energy and/or

repeat discharges as
needed.

• Consider replacing

disposable therapy
electrodes with new.

9

CONNECT ELECTRODES

message appears.

Inadequate connection to
defibrillator.

• Check for electrode

connection.

Electrodes do not adhere
properly to the patient.

• Press electrodes firmly on

patient’s skin.

• Clean, shave, and dry the

patient’s skin as
recommended.

Electrodes are dry, damaged,
or out of date.

• Replace the electrodes.

10

CONNECT TO AC POWER

message appears.

Remote sync is selected and
the defibrillator is not
connected to AC power.

• Connect to AC power.
• Press

SYNC

to turn off remote

sync.

11

CONNECT SYNC CABLE TO

REMOTE MONITOR

message

appears.

Remote sync is selected and
the defibrillator is not
connected to the remote
monitor.

• Connect to remote monitor.
• Press

SYNC

to turn off remote

synchronization.

12

LOW IMPEDANCE–

RECHARGING

message

appears.

Patient impedance of <15
ohms detected.

• No action required.

13

SEARCHING FOR SIGNAL

message appears.

Remote sync is selected and
the defibrillator is qualifying
the input signal.

• No action required.

Table 4-3

Troubleshooting Tips for Defibrillation and Synchronized Cardioversion (Continued)

Observation

Possible Cause

Corrective Action