Physio-Control LIFEPAK 20e User Manual

Page 156

Advertising
background image

Specifications and Performance Characteristics

A-4

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

CO

2

CO

2

Range

0 to 99 mmHg (0 to 13.2 kPa)
Units: mmHg, %, or kPa

CO

2

Accuracy

CO

2

partial pressure at sea

level:

Accuracy:

(0–80 bpm)*

0 to 38 mmHg
(0 to 5.1 kPa)

±2 mmHg

(0.27 kPa)

39 to 99 mmHg
(5.2 to 13.2 kPa)

±5% of reading + 0.8% for
every 1 mmHg (0.13 kPa)
above 38 mmHg (5.1 kPa)

(>80 bpm)*

0 to 18 mmHg
(0 to 2.4 kPa)

±2 mmHg
(0.27 kPa)

19 to 99 mmHg
(2.53 to 13.2 kPa)

±4 mmHg (0.54 kPa) or
±12% of reading, whichever
is higher

*For RR > 60 bpm, to achieve specified CO

2

accuracy, the

Microstream

®

Filterline

®

H Set for infants must be used.

Drift of Measurement
Accuracy

No drift in accuracy for at least 6 hours

Respiration Rate
Accuracy

0 to 70 bpm: ±1 bpm
71 to 99 bpm: ±2 bpm

Respiration Rate
Range

0 to 99 breaths/minute

Flow Rate

42.5 to 65 ml/min (measured by volume)

Rise Time

190 msec

Response Time

4.5 seconds, maximum (includes delay time and rise time)

Initialization Time

30 seconds (typical), 10–180 seconds

Ambient Pressure

Automatically compensated internally

Waveform Scale
Factors

Autoscale, 0–20 mmHg (0–4 Vol%), 0–50 mmHg (0–7 Vol%),
0–100 mmHg (0–14 Vol%)

PRINTER

Prints continuous strips of the displayed patient information

Paper Size

50 mm (2.0 in.)

Print Speed

Continuous ECG 25 mm/sec

±

5% (measured in accordance

with AAMI EC-11 (1991), 4.2.5.2)

25 mm/sec print speed for CODE SUMMARY Reports

Delay

8 seconds

Autoprint

Waveform events print automatically (user configurable)