LIFEPAK 20e Defibrillator/Monitor Operating Instructions

4-9

©2006-2013 Physio-Control, Inc.

Proper Connection

After a

NO SHOCK ADVISED

prompt you

will see and hear

START CPR

. A

countdown timer (min:sec format)
continues for the duration specified in
the

CPR TIME 2

setup option. (Refer to

Section 8

).

Note: If the CPR metronome is set
to be on during CPR times, you will
hear audible “tocks” and ventilation
prompts or tones.

When the CPR countdown time ends,
you will see and hear

PUSH ANALYZE

.

This message stays on the screen and
the voice prompt will repeat every 20
seconds until you press the

ANALYZE

button.

If therapy electrodes are not connected
to the therapy cable or not placed on the
patient’s chest, the

CONNECT

ELECTRODES

message and voice

prompt occur until the patient is
connected to the AED.

If the therapy cable is not connected to
the defibrillator, you will see the

CONNECT CABLE

message until the

cable is connected.

If the test plug is connected to the
therapy cable when analysis is initiated,
the

REMOVE TEST PLUG

message and

voice prompt occurs.

After you remove the test plug from the
therapy cable, the message and voice
prompt

CONNECT ELECTRODES

occurs

until the patient is connected to the
AED.