Physio-Control LIFEPAK 20e User Manual

Page 78

Advertising
background image

Therapy

4-8

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

No Shock Advised

Note: If you do not press the

button within 60 seconds, the AED
disarms the shock button, and the

DISARMING

message appears.

When the

button is pressed you will

see the message

ENERGY DELIVERED

indicating energy transfer was
completed.

When energy transfer is complete the
shock counter increases by 1. This will
continue to increase incrementally with
every energy transfer.

After a shock is delivered you will see
and hear

START CPR

. A countdown

timer (min:sec format) continues for the
duration specified in the

CPR TIME 1

setup option.

Note: If the CPR metronome is set
to be on during CPR times, you will
hear audible “tocks” and ventilation
prompts or tones.

When the CPR countdown time ends,
you will see and hear

PUSH ANALYZE

.

This message stays on the screen and
the voice prompt will repeat every 20
seconds until you press the

ANALYZE

button.

If the AED detects a nonshockable
rhythm, you will see and hear

NO

SHOCK ADVISED

. The AED will not

charge, and a shock can not be
delivered.

Advertising