Special aed setup options, Special aed setup options -10 – Physio-Control LIFEPAK 20e User Manual

Page 80

Advertising
background image

Therapy

4-10

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

Motion Detected

Continuous Patient Surveillance System

Special AED Setup Options

The following descriptions of voice prompts and messages explain special cprMAX technology
setup options (refer to

Appendix F

).

Initial CPR - CPR First

When the

INITIAL CPR

option is set to

CPR FIRST

, you will be prompted to

START CPR

immediately after the AED is turned on.

If motion is detected during the ECG
analysis, you will see and hear

MOTION

DETECTED, STOP MOTION

followed by a

warning tone. Analysis is inhibited up to
10 seconds. After 10 seconds, even if
motion is still present, the analysis
proceeds to completion. Refer to

Table

4-1

for possible motion causes and

suggested solutions.

When the device is not analyzing the
ECG and is not in CPR Time, the
Continuous Patient Surveillance System
(CPSS) actively monitors the ECG for a
potentially shockable rhythm.
If a shockable rhythm is detected, you
will see and hear

PUSH ANALYZE

.

You should:
1 Confirm the patient is unconscious,

pulseless, and not breathing
normally.

2 Confirm no motion is present. Stop

CPR.

3 Press

ANALYZE

.

The AED begins to analyze the patient’s
ECG.
For information about changing the
defibrillation mode, refer to

Section 8

.

You will see and hear

START CPR

.