Screen Messages

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

C-5

©2006-2013 Physio-Control, Inc.

STAND CLEAR/PUSH
SHOCK BUTTON

Stand clear of the patient and push the

SHOCK

button.

START CPR

Initiate CPR in AED mode.

SWITCHING PRIMARY
TO LEAD II

PADDLES

lead is not available and you pressed the

ANALYZE

button.

SWITCHING PRIMARY
TO PADDLES

PADDLES

lead is available and you pressed the

ANALYZE

button.

TRANSMISSION
CANCELLED

Data transmission has been cancelled.

TRANSMISSION
COMPLETE

Data transmission completed successfully.

TRANSMISSION
FAILED

Data transmission was not successful.

UNABLE TO TRANSMIT

Unable to send data.

USE ECG LEADS

The device is attempting synchronized cardioversion and

PADDLES

lead was selected.

USER TEST DID NOT
COMPLETE

Test plug not connected to QUIK-COMBO therapy cable, standard
paddles not seated in paddle wells, possible defective therapy cable,
or a problem with the defibrillator during user test.

USER TEST DID NOT
COMPLETE –
CONNECT TO TEST
PLUG

Test plug not connected to QUIK-COMBO therapy cable, standard
paddles not seated in paddle wells, defective therapy cable, or a
problem with the defibrillator during user test.

USER TEST FAILED

An unsuccessful user test has occurred.

USER TEST IN
PROGRESS

Confirms that the user test is in progress.

USER TEST
SUCCEEDED

A successful user test was completed.

XX% TRANSMITTED

Specified percent of the transmission is completed.

Table C-1

Summary of Screen Messages (Continued)