Presented in, Table b-1, Table b-2 – Physio-Control LIFEPAK 20e User Manual

Page 170

Advertising
background image

Clinical Summaries

B-4

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

Cumulative percentages of successes for cardioversion of AF with shocks of 200 J or less, the
primary endpoint of the study, was significantly higher in the biphasic group than the monophasic
group (p<0.0001). The observed cumulative percentage of successes at 360 J was also higher
for biphasic shocks than for monophasic shocks, but did not attain statistical significance.

Table B-1

Cumulative Success Rates and Crossover Results for Cardioversion of AF

Energy Setting

70 J

100 J

200 J

360 J

360 J Crossover Successes

MDS: n = 37

5.4%

19%

38%

86%

4 of 5 pts succeeded with
360 J BTE shock

BTE: n = 35

60%

80%

97%

97%

0 of 1 pts succeeded with
360 J MDS shock

Table B-2

Energy Settings, Delivered Energy and Peak Current for Shocks Delivered to Patients in AF

Energy Setting

Number of Patients

Delivered Energy

Peak Current, Amps

Monophasic shocks

70 J

37

73±3

21.0±3.5

100 J

35

105±4

24.6±4.3

200 J

30

209±7

34.6±5.9

360 J

23

376±13

46.8±8

360 J crossover
shocks

1

380

44.7

Biphasic shocks*

70 J

35

71±0

11.9±2.5

100 J

14

102±0

14.9 ±3.5

200 J

7

203±1

20.6±3.5

360 J

1

362

28.5

360 J crossover
shocks

5

361±6

32.4±8.5

* Peak current and delivered energy are not available for two of the patients treated with biphasic shocks.