Physio-Control LIFEPAK 20e User Manual

Page 61

Advertising
background image

3

M

o

n

itoring

Monitoring

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

3-15

©2006-2013 Physio-Control, Inc.

6

SPO2: UNKNOWN SENSOR

message appears.

A sensor is connected to the
device that is not compatible
with the Masimo Sp0

2

module.

• Refer to

"Pulse Oximeter

Sensors," page 3-13

for

sensor compatibility.

7

SPO2: SEARCHING FOR
PULSE

message appears.

A sensor is connected to the
patient and is searching for a
pulse.

• Wait for completion.

8

SPO2: LOW PERFUSION

message appears.

Patient has a weak pulse.

• Change sensor location.

Table 3-3

Troubleshooting Tips for SpO

2

(Continued)

Observation

Possible Cause

Corrective Action

Advertising