Therapy

4-22

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

Defibrillation Procedure

To defibrillate the patient:

1

Press

ON

to turn on the defibrillator.

2 To access the pediatric paddles, slide the adult paddle forward until it releases.

3 Apply defibrillation gel to the pediatric paddle electrode surfaces.

4 Select the appropriate energy for the weight of the child according to American Heart

Association recommendations (or equivalent guidelines).

5 Place the paddles firmly on the patient's chest.

6 Press

CHARGE

.

7 Make certain all personnel, including the operator, are clear of the patient, the bed, and any

equipment connected to the patient.

8 Confirm ECG rhythm and available energy.

9 Press

the

SHOCK

button on the defibrillator or the

SHOCK

buttons on the standard paddles to

discharge energy to the patient. For standard paddles, apply firm pressure with both paddles to
the patient’s chest, and press both paddle buttons simultaneously to discharge energy to the
patient. For safety reasons, the

SHOCK

button on the defibrillator front panel is disabled when

using standard paddles.

Note: To disarm (cancel the charge), press the Speed Dial. The defibrillator disarms

automatically if shock buttons are not pressed within 60 seconds, or if you change the
energy selection after charging begins. Press

CHARGE

to restart charging.

Note: If the

ABNORMAL ENERGY DELIVERY

message appears and the shock is not effective,

increase energy, if necessary, and repeat shock. (Also refer to

page 4-24

.)

Troubleshooting Tips for Defibrillation and Synchronized Cardioversion

Table 4-3

Troubleshooting Tips for Defibrillation and Synchronized Cardioversion

Observation

Possible Cause

Corrective Action

1 Charge time to 360 J

exceeds 10 seconds.

Battery low.

• Connect to AC power.
• Defibrillator operating in low

temperature environment
(<25°C or 77°F).

2 Energy not delivered to

patient when

SHOCK

button(s) are pressed.

Defibrillator in sync mode and
QRS complexes are not
detected.

• Change ECG lead for

optimum sensing QRS or
deactivate

SYNC

.

Defibrillator in sync mode;

SHOCK

button(s) not pressed

and held until next detected
QRS.

• Hold shock buttons until

discharge occurs or next
detected QRS.

SHOCK

button(s) pressed

before full charge reached.

• Wait for tone and message

indicating full charge.

Sixty seconds elapsed before

SHOCK

button(s) pressed

after full charge. Energy
internally removed.

• Press

SHOCK

button(s) within

60 seconds of full charge.

ENERGY

selection changed.

• Press

CHARGE

again.

Test plug connected to
QUIK-COMBO therapy cable.

• Disconnect test plug and

connect electrodes to
QUIK-COMBO therapy cable.