Troubleshooting Tips for AED Mode

Table 4-1

Troubleshooting Tips for AED Mode

Observation

Possible Cause

Corrective Action

1

CONNECT ELECTRODES

message appears.

Inadequate connection to
defibrillator.

• Check for electrode

connection.

Electrodes do not adhere
properly to the patient.

• Press electrodes firmly on

patient’s skin.

• Clean, shave, and dry the

patient’s skin as
recommended.

Electrodes are dry, damaged,
or out of date.

• Replace the electrodes.

2

REMOVE TEST PLUG

message appears.

Test plug connected to therapy
cable.

• Disconnect test plug from

therapy cable.

• Connect electrodes to

therapy cable.

3

MOTION DETECTED

and

STOP MOTION

messages

appear during analysis.

Patient movement.

• Stop CPR during analysis.
• When patient is being

manually ventilated, press

ANALYZE

after complete

exhalation

.

• Move patient to stable

location when possible.

Patient movement because of
agonal respirations.

• Press

ANALYZE

immediately

after exhalation or wait until
agonal respirations are
slower or absent.

Electrical/radio frequency
interference.

• Move hand-held

communication devices or
other suspected devices
away from the defibrillator
when possible.

4

DISARMING

message

appears.

Therapy cable or electrodes
disconnected from AED or
patient.

• Confirm secure connection

and press

ANALYZE

.

Shock button not pressed
within 60 seconds or door is
open.

• Press

ANALYZE

again.

• Press

SHOCK

immediately

when directed.

5 Voice prompts sound faint

or distorted.

Low battery power.

• Connect to AC power.

6

LOW IMPEDANCE–
RECHARGING

message

appears.

Patient impedance <15 ohms
detected.

• No action required.

7 Metronome not heard.

Metronome only sounds during
CPR time.

• No action required.

Metronome not set to be

ON

in

AED mode.

• Refer to

"AED Mode Setup

Menu," page 8-7

.