Physio-Control LIFEPAK 20e User Manual

Page 69

Advertising
background image

3

M

o

n

itoring

Monitoring

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

3-23

©2006-2013 Physio-Control, Inc.

8 EtCO

2

values are

consistently higher than
expected.

Physiological cause such as
COPD.

• None.

Inadequate ventilation.

• Check ventilator; increase

ventilatory rate/bagging.

Patient splinting during
breathing.

• Supporting measures such

as pain relief.

Improper calibration.

• Contact qualified service

personnel.

9 EtCO

2

values are

consistently lower than
expected.

FilterLine connection to device
is loose.

• Twist FilterLine connector

clockwise until firmly seated
and can no longer be
turned.

Physiological cause.

• See Physiological factors in

"CO

2

Detection," page 3-21

.

Hyperventilation.

• Check ventilator; decrease

ventilatory rate/bagging.

Improper calibration.

• Contact qualified service

personnel.

10 CO

2

waveform stays

elevated for several
seconds.

Expiration is prolonged due to
bagging technique.

• Release bag reservoir

completely with expiration.
Observe that elevated
baseline returns to normal
level.

11 Sudden extreme increase in

EtCO

2

.

Fluid has entered CO

2

module. • Contact qualified service

personnel.

12

CHECK CO2

message

appears.

FilterLine blockage

• Disconnect and then

reconnect the FilterLine set.

• Change the FilterLine set.

Exhaust port blockage

• Ensure exhaust port on the

back of the device is not
blocked.

• Check exhaust tubing (if

present) for blockages or
kinks.

RF equipment too close to
defibrillator

• Separate RF equipment

from defibrillator. See
recommended separation
distances in

Table 4, page

H-4

.

13

XXX

appears instead of

EtCO

2

value.

CO

2

module malfunction.

• Turn device off and then on

again.

• If problem persists, contact

qualified service personnel.

14 There is no EtCO

2

value

and the CO

2

waveform is

flat.

Measured CO

2

is less than

3.5 mmHg.

• See

"CO

2

Detection,"

page 3-21

.

Table 3-4

Troubleshooting Tips for EtCO

2

(Continued)

Observation

Possible Cause

Corrective Action