LIFEPAK 20e Defibrillator/Monitor Operating Instructions

F-3

©2006-2013 Physio-Control, Inc.

Stacked Shocks

When set to

OFF

, the

STACKED SHOCKS

option inserts prompting for CPR after each (a single)

shock. This eliminates the three-shock stack. CPR is prompted after the shock regardless of the
ECG rhythm. The CPR time following the shock is determined by the

CPR TIME 1

setting

selected. Choices for the

STACKED SHOCKS

option are

ON

or

OFF

. The default setting is

OFF

.

When this option is set to

ON

, the defibrillator follows the previously traditional stacked shock

protocol and delivers up to three consecutive shocks, as necessary, without interposed CPR.

Pulse Check

The

PULSE CHECK

option inserts prompting to check for a pulse or check the patient, depending

on the

PULSE PROMPT

setting. The choices for

PULSE CHECKS

are:

ALWAYS, AFTER EVERY

NSA, AFTER SECOND NSA

, and

NEVER

. The default setting is

NEVER

.

• The

ALWAYS

option prompts for a pulse check after CPR Times 1 and 2, after a

NO SHOCK

ADVISED

decision, after a single

SHOCK ADVISED

decision with

STACKED SHOCKS OFF

, or

after three consecutive

SHOCK ADVISED

decisions if

STACKED SHOCKS

is

ON

.

• The

AFTER EVERY NSA

option prompts for a pulse check after every

NO SHOCK ADVISED

decision.

• The

AFTER SECOND NSA

option prompts for a pulse check after the second analysis if the

second analysis results in a

NO SHOCK ADVISED

decision, regardless of the first analysis

decision (

SHOCK ADVISED

or

NO SHOCK ADVISED

).

• The

NEVER

option eliminates all

PULSE CHECK

prompts.