Table 2-2

Battery Status Indicator

Battery

Indicator

Battery Status

Messages and Tones

Battery

Capacity

(percentage)

Four green bars

None

76–100

Three green bars

None

51–75

Two green bars

None

26–50

One green bar

None

16–25

One yellow bar

LOW BATTERY

Three beeps (one time). Message
remains in message area, cycling with
others if necessary.

11–15

One flashing red bar

LOW BATTERY: CONNECT TO AC POWER

Three beeps (every 20 seconds).
Message remains in message area,
cycling with others if necessary.

0–10

It is important to note that both charge (voltage) and capacity are factors contributing to the
battery status indicator’s display. For example, the battery status indicator displays two green
bars when a new battery has 100% capacity and has been charged to 50%. However, an
older battery that is 100% charged and only has 50% capacity also displays two green bars.
The following demonstrates this example.

This battery may either be low in capacity or low in charge as follows:

• For a new battery with 100% capacity, this symbol indicates 50% charged (low in

charge).

• For an older battery with 50% capacity, this symbol indicates 100% charged (low

in capacity).