Specifications and Performance Characteristics

A-8

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

PACER

Pacing Mode

Demand or non-demand
Rate and current defaults (user configurable)

Pacing Rate

40 to 170 ppm

Rate Accuracy

±

1.5% over entire range

Output Waveform

Monophasic, amplitude stable to ±5% relative to leading edge
for currents greater than or equal to 40 mA, Duration 20
±1 msec, Rise/Fall times <= 1 msec [10-90% levels]

Output Current

0 to 200 mA

Pause

Pacing pulse frequency reduced by a factor of 4 when
activated

Refractory Period

200 to 300 msec

±

3% (function of rate)

ENVIRONMENTAL

Temperature, Operating

5° to 40°C (41° to 104°F)

Temperature, Non-
Operating

-20° to 60°C (-4° to 140°F) except therapy electrodes

Relative Humidity,
Operating

5 to 95%, non-condensing

Atmospheric Pressure,
Operating

Ambient to 522 mmHg (0 to 3048 meters) (0 to 10,000 ft)

Water Resistance,
Operating (without
accessories except for
ECG Cable and Hard
Paddles)

IPX1 (spillage) per IEC 60601-1 clause 44.6

EMC

IEC 60601-1-2:2001/EN 60601-1-2:2001, Medical Equipment-
General Requirements for Safety-Collateral Standard:
Electromagnetic Compatibility-Requirements and Tests.
IEC 60601-2-4:2002; Clause 36/EN 60601-2-4:2003: Clause
36, Particular Requirements for the Safety of Cardiac
Defibrillators and Cardiac Defibrillator monitors.

Shock (drop)

1 drop on each side from 45.7 cm (18 in.) onto a steel surface

Vibration

MIL-STD-810E Method 514.4, Cat 1