GENERAL TROUBLESHOOTING TIPS

If a problem with the defibrillator/monitor is detected during operation or testing, refer to the
troubleshooting tips in

Table 7-2

. If the problem cannot be corrected, remove the defibrillator/

monitor from use and contact qualified service personnel.

Table 7-2

General Troubleshooting Tips

Observation

Possible Cause

Corrective Action

1 No power when

defibrillator/monitor is
turned

ON.

Low battery voltage.

• Connect to AC power.

2 Defibrillator/monitor

operates, but screen is
blank.

Operating temperature is
too low or too high.

• Connect to AC power.

Screen is not displaying
properly.

• Contact qualified service personnel.

3

CHECK PRINTER

message appears.

Printer paper jams, slips, or

misfeeds.

Printer is out of paper.

• Reinstall paper.

• Add new paper.
• If problem persists, contact qualified

service

personnel

.

4 AC Mains LED on

defibrillator/monitor not
illuminated when
connected to AC power.

Loose or improper
connection between
defibrillator and power
source.

• Check power connections and

cables.

Loose or improper
connection between
defibrillator and
CodeManagement Module.

• Contact qualified service personnel.

5 Service LED

illuminates.

Device self-test circuitry
detects service condition.

• Continue to use defibrillator or

pacemaker if needed.

• Turn device off and then on again.

Note that this creates a new Patient
Record.

• If service LED does not clear,

remove device from active use.

• Report occurrence of service LED to

qualified service personnel.

6 ECG monitoring

problems.

• Refer to

Section 3

,

page 3-5

.

7 Problems with AED

operation.

• Refer to

Section 4

,

page 4-13

.

8 Problems with

defibrillation/
synchronized
cardioversion.

• Refer to

Section 4

,

page 4-22

.

9 Problems with pacing.

• Refer to

Section 4

,

page 4-25

.

10 Displayed time is

incorrect.

Time is incorrectly set.

• Change the time setting. Refer to

Section 2

,

page 2-7

.

Advertising