Physio-Control LIFEPAK 20e User Manual

Page 133

Advertising
background image

7

M

ainta

ining the

Equipment

Maintaining the Equipment

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

7-11

©2006-2013 Physio-Control, Inc.

11 Date printed on report

is incorrect.

Date is incorrectly set.

• Change the date setting. Refer to

Section 2

,

page 2-7

.

12 Displayed messages

are faint or flicker.

Low battery power.
Out of temperature range.

• Connect to AC power immediately.

13 Low speaker volume.

Moisture in speaker grill
holes.

• Wipe moisture from speaker grill and

allow device to dry.

14

MAINTENANCE DUE

message appears.

Maintenance prompt is set
to display at a selected
interval in service mode.

• Continue to use device if needed.
• Contact service personnel to reset or

turn off the maintenance prompt.

• Contact Physio-Control Technical

Support for instructions on how to
reset or turn off this prompt.

15

SELF TEST DID NOT

COMPLETE

message

appears.

Test plug not connected to
QUIK-COMBO therapy
cable during daily auto test.

• Connect test plug to QUIK-COMBO

therapy cable.

• Perform manual user test.

Standard paddles not
seated in paddle wells
during daily auto test.

• Make sure standard paddles are

securely seated in the paddle wells
for daily auto test.

Defective therapy cable or
a problem with the
defibrillator.

• Contact qualified service personnel.

16

SELF TEST FAILED

message appears.

Device self-test circuitry
detects service condition
during daily auto test.

• Use defibrillator or pacemaker if

needed in an emergency.

• Report occurrence of self-test failure

to qualified service personnel.

17

USER TEST DID NOT

COMPLETE

message

appears.

Test plug not connected to
QUIK-COMBO therapy
cable during user test.

• Connect test plug to QUIK-COMBO

therapy cable for user test.

Standard paddles not
seated in paddle wells
during user test.

• Make sure standard paddles are

securely seated in the paddle wells
for user test.

Defective therapy cable or
a problem with the
defibrillator.

• Contact qualified service personnel.

18

USER TEST FAILED

message appears.

Device self-test circuitry
detects service condition
during user test.

• Use defibrillator or pacemaker if

needed in an emergency.

• Report occurrence of user-test

failure to qualified service personnel.

19 Printout annotated

79 J, but > 79 J was
selected.

Paddles discharged when
seated in the paddle wells
or shorted together.

• Refer to

"Standard Paddles

Defibrillation and Synchronized
Cardioversion Check on Battery
Power," page 7-6

.

• Perform User Test.

20 Defibrillator does not

charge to 200 J within
10 seconds while on
battery power.

Low battery condition.

Defective battery.

• Connect the defibrillator to AC

power.

• Contact qualified service personnel.

Table 7-2

General Troubleshooting Tips (Continued)

Observation

Possible Cause

Corrective Action