Co2 display, Co2 alarms, Display – Physio-Control LIFEPAK 20e User Manual

Page 66: Alarms

Advertising
background image

Monitoring

3-20

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

CO

2

Display

The following scales are available to display the CO

2

waveform. The LIFEPAK 20e

defibrillator/monitor automatically selects the scale based on the measured EtCO

2

value. To

change the CO

2

scale, outline and select the CO

2

area using the Speed Dial

and then select the

desired scale from the scale menu.

• Autoscale (default)

• 0–20 mmHg (0–4 Vol% or kPa)

• 0–50 mmHg (0–7 Vol% or kPa)

• 0–100 mmHg (0–14 Vol% or kPa)

The CO

2

waveform is compressed (displayed at 12.5 mm/sec sweep speed) to provide more data

in the 3.7-second screen. There is a slight delay between when the breath occurs and when it
appears on the screen. Printouts are at 25 mm/sec.

The monitor shows the maximum CO

2

value over the last 20 seconds. If the EtCO

2

values are

increasing, the change can be seen with every breath. However, if the values are continually
decreasing, it will take up to 20 seconds for a lower numerical value to be displayed. Because of
this, the EtCO

2

value may not always match the level of the CO

2

waveform.

CO

2

Alarms

The EtCO

2

monitor provides:

• EtCO

2

high and low alarms (controlled by activating Alarms (refer to

"Setting Alarms,"

page 2-17

)).

• FiCO

2

(inspired CO

2

) alarm (controlled by activating Alarms; automatically set to default values

and not adjustable)

• Apnea alarm (automatic and not adjustable)

• Respiration Rate high and low alarms (controlled by activating Alarms (refer to

"Setting

Alarms," page 2-17

)).

Note: The apnea alarm occurs when a breath has not been detected for 30 seconds. The
message

ALARM APNEA

appears in the message area along with the time since the last

detected breath.

Advertising