Replacing the adult paddle attachment, Cleaning the standard paddle set, Figure 5-9 – Physio-Control LIFEPAK 20e User Manual

Page 106

Advertising
background image

Paddle Accessory Options

5-8

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

4 The pediatric paddle is now exposed and ready for use (refer to

Figure 5-10

).

Replacing the Adult Paddle Attachment

To replace the adult paddle attachment:

1 Hold the adult paddle attachment with one hand and the standard handle with the other hand.

2 Fit the pediatric paddle onto the adult paddle attachment.

3 Slide the paddle handle forward until you hear a click. (

Refer to

Figure 5-11

.)

Figure 5-11

Replacing a Pediatric Paddle

Each adult paddle attachment has a contact spring plate that transfers energy from the pediatric
paddle to the adult paddle. Routinely inspect the spring plates and pediatric paddle surfaces to
make sure that they are clean and intact.

Cleaning the Standard Paddle Set

Individually protect paddles before and after cleaning to prevent damage to paddle surfaces.
After each use:

1 Separate the adult and pediatric paddles.

2 Wipe or rinse paddle electrodes, cable connector, handles, and cables with mild soap and

water or disinfectant using a damp sponge, towel, or brush. Do not immerse or soak.

3 Dry all parts thoroughly.

4 Examine paddles (including electrode surfaces), cables, and connectors for damage or signs

of wear.

Note: Cables showing signs of wear, such as loose cable connections, exposed wires, or

cable connector corrosion, should be removed from use immediately.

Note: Paddles with rough or pitted electrodes should be removed from use immediately.

Figure 5-9

Accessing a Pediatric Paddle

Figure 5-10

Pediatric Paddle (Bottom)

Pediatric
paddle

Adult paddle
attachment

Pediatric
paddle

Adult paddle
attachment