Basic Orientation

2-18

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

When the alarms are set to

OFF

, you must press the

ALARMS

button and select

QUICK SET

to

enable the alarms.

When you press the

ALARMS

button, the following Alarms

overlay appears:

1 Select

QUICK SET

to activate the

alarms for all active parameters. The
quick set limits are set automatically
based on the patient’s current vital
sign values (refer to

Table 2-1

). The

alarm limits default to the setting
(

WIDE

or

NARROW

) displayed on the

overlay.

2 Select

LIMITS

to change the alarm

limits to

WIDE

or

NARROW

(refer to

Table 2-1

).

3 Select

SUSPEND

to turn off the

audible alarm for up to 15 minutes. If
an alarm limit is exceeded while the
alarm is silenced, the violated
parameter flashes, an alarm
message appears, but the alarm tone
remains silent.

Select

VF/VT ALARM

to turn on

continuous monitoring for ventricular
fibrillation and ventricular tachycardia
in manual mode.

A symbol appears above the primary

ECG when the alarm is on

.

Reselect

VF/VT

ALARM

to turn off

this alarm.

Note: When the VF/VT alarm is

on, you are limited to

PADDLES

lead or lead II. Refer to

"Selecting

ECG Lead and Size," page 3-2

.

Note: The VF/VT alarm will be

suspended when the noninvasive
pacemaker is on and when
standard paddles are attached
and

PADDLES

lead is selected.

The alarm is also suspended
when the defibrillator is charging
or is fully charged.

Advertising