Electromagnetic Compatibility Guidance

H-4

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

Table 4

Recommended Separation Distances between Portable and Mobile RF Communications

Equipment and the LIFEPAK 20e Defibrillator/Monitor Series

The LIFEPAK 20e defibrillator/monitor is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the LIFEPAK 20e defibrillator/monitor can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the LIFEPAK 20e defibrillator/monitor as recommended below,
according to the maximum output power of the communications equipment.

Rated maximum

output power of

transmitter

W

Separation distance according to frequency of transmitter

m

150 kHz to 80 MHz

outside ISM bands

d = 1.2

P

150 kHz to 80 MHz

in ISM bands

d = 1.2

P

80 MHz to 800 MHz

d = 1.2

P

800 MHz to 2.5 GHz

d = 2.3

P

0.01

0.12

0.12

0.12

0.23

0.1

0.38

0.38

0.38

0.73

1

1.2

1.2

1.2

2.3

10

3.8

3.8

3.8

7.3

100

12

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

Note 2: The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to

6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.

Note 3: An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters

in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5
GHz to decrease the likelihood that mobile/portable communications equipment could cause
interference if it is inadvertently brought into patient areas.

Note 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption

and reflection from structures, objects and people.