Setup Options

8-2

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

SETUP OPTIONS

Setup options allow you to define operating features for the LIFEPAK 20e defibrillator/monitor
such as device identification numbers and default settings.

Table 8-1

through

Table 8-21

list all

setup options along with the factory default settings.

Setup options can be selected in either of two ways:

• Use the Setup Options menu on the LIFEPAK 20e device. See

"Entering Setup Options,"

page 8-3

, for more information.

• Use LIFENET Device Agent to manage setup options on the LIFEPAK 20e device. See the

Help system for LIFENET Device Agent for more information.

Print Configurations Before Service or Repair

If the defibrillator receives service or repair that affects the internal memory components, such as
replacement of the main printed circuit board, any changes previously made to the option
definitions may be lost from memory. Before allowing service or repair, be sure to print the
current user setup defaults so the customized definitions may be reentered after service or
repair. (Refer to

"Print Defaults," page 8-13

.)

Passcode Security

To prevent unauthorized access, a security passcode is required for access to the Setup menu
and service mode (refer to

page 8-14

). The LIFEPAK 20e defibrillator/monitor allows you to

change both of these passcodes. The passcode definition is part of the device identification
option.

Note: To use the defibrillator with any new settings, you must turn the defibrillator off and on
again.

WARNING!

Possible improper defibrillator performance.

Changing factory default settings will change the behavior of the device. Changes to the default
settings must only be made by authorized personnel.

WARNING!

Possible ineffective energy delivery.

Service mode is for authorized personnel only. Improper use of service mode may inappropriately
alter the defibrillator’s configuration and may change energy output levels. Contact service
personnel for assistance or information about defibrillator configuration.