3

M

o

n

itoring

Monitoring

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

3-3

©2006-2013 Physio-Control, Inc.

You can select or change the ECG size by using the

SIZE

button or the Speed Dial. If an ECG is

in Channel 2, the size is automatically changed to match the Channel 1 size.

To select or change the ECG size using the Speed Dial:

Adjusting the Systole Tone Volume

To adjust the systole tone volume, highlight and select heart rate (HR) in the monitoring area of
the screen.

The following overlay appears:

To select or change the ECG lead using
the Speed Dial:
1 Highlight and select

CHANNEL 1

, and

then select

LEAD

to obtain the

primary ECG lead choices.

2 Change the ECG lead by rotating the

Speed Dial. The highlighted selection
shows the ECG lead.

3 Repeat Step 1 and Step 2 to select or

change displayed waveforms for
Channel 2.

To select or change the ECG size using
the

SIZE

button:

1 Press

the

SIZE

button.

2 When the Size menu appears, press

the

SIZE

button again. The

highlighted selection shows the
current ECG size.

1 To obtain the primary ECG, highlight

and select

CHANNEL 1

, and then

select

SIZE

.

2 Change ECG size by rotating the

Speed Dial. The highlighted selection
shows the current ECG size.

1 Rotate the Speed Dial to the desired

volume.

2 Press the home screen to exit.