Therapy

4-6

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

When the CPR metronome is set to

ON

in the setup options (refer to

Section 8

), the CPR

metronome icon

is displayed and the metronome sounds automatically during CPR times.

You cannot enable or disable the CPR metronome while in AED mode. Refer to

"CPR

Metronome," page 4-17

, for more information.

AED Procedure

The following descriptions of voice prompts and messages are based on the factory default
settings for AED mode. The default settings are consistent with 2010 American Heart
Association (AHA) and European Resuscitation Council (ERC) guidelines. Changing the setup
options may result in different AED behavior. Refer to

Section 8

for setup option choices.

1 Verify that the patient is in cardiopulmonary arrest (unresponsive, not breathing normally, and

showing no signs of circulation.)

2 Press

ON

.

3 Prepare the patient for electrode placement (refer to

"Paddles Monitoring Procedure"

on

page 3-4

).

4 Connect the therapy electrodes to the therapy cable, and confirm cable connection to the

defibrillator.

5 Apply the electrodes to the patient's chest in the anterior-lateral position (refer to

"Anterior-

lateral Placement," page 4-3

).

When the ECG waveform is set to

OFF

in

the setup options, the messages and
prompts fill the screen as shown in the
screen to the left.

The

CONNECT ELECTRODES

message

and voice prompt occur until the patient
is connected to the AED.