Defibrillation procedure, Defibrillation procedure -16 – Physio-Control LIFEPAK 20e User Manual

Page 86

Advertising
background image

Therapy

4-16

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

If the defibrillator is charged to 80 joules or more with paddles in the paddle wells, and then the
paddles are removed and placed on a patient, the defibrillator continues charging to the selected
energy and defibrillation may be completed as usual. When charging the defibrillator with the
paddles on the patient’s chest, the defibrillator automatically adjusts the waveform voltage and
current duration based on the patient's impedance. It is important to understand that this
automatic adjustment does not occur when the defibrillator is charged with the paddles seated in
the paddle wells.

When the paddles or QUIK-COMBO electrodes are on the patient's chest when the defibrillator is
charged and the defibrillator measures an impedance of 15 ohms or less, the defibrillator disarms
the capacitor and automatically recharges to a lower energy setting. When this condition occurs,
the

LOW IMPEDANCE-RECHARGING

message appears on the display. When charging is

complete, defibrillation may be completed as usual.

Defibrillation Procedure

You can configure the LIFEPAK 20e defibrillator/monitor to automatically sequence energy
levels. Refer to

"Manual Mode Setup Menu," page 8-5

.

1 Press

ON

.

2 Identify the electrode or paddle sites on the patient. Use either the anterior-lateral or anterior-

posterior position as described on

page 4-3

.

3 Prepare the patient’s skin for electrode application:

• If possible, place the patient on a firm surface away from standing water.
• Remove clothing from the patient’s upper torso.
• Remove excessive hair from the electrode sites; if shaving is necessary, avoid cutting the

skin.

• Clean the skin and dry it briskly with a towel or gauze.
• Do not apply alcohol, tincture of benzoin, or antiperspirant to the skin.

4 Connect the therapy electrodes to the therapy cable, and confirm cable connection to the

defibrillator.

Note: When in

PADDLES

lead, if the

REMOVE TEST PLUG

message appears, disconnect

test plug and connect therapy electrodes to QUIK-COMBO therapy cable.

5 Apply therapy electrodes to the patient in anterior-lateral or anterior-posterior position. If using

standard paddles, apply conductive gel to the paddles and place paddles on the patient’s
chest.

6 Press

ENERGY SELECT

.

7 Press

CHARGE.

While the defibrillator is charging, a charging bar appears and a ramping tone

sounds, indicating the charging energy level. When the defibrillator is fully charged, an overlay
appears (refer to

"Defibrillation Procedure," page 4-16

).

8 Make certain all personnel, including the operator, stand clear of the patient, bed, and any

equipment connected to the patient.

9 Confirm ECG rhythms and available energy.

10 Press the

SHOCK

button on the defibrillator or the

SHOCK

buttons on the standard paddles to

discharge energy to the patient. For standard paddles, apply firm pressure with both paddles to
the patient’s chest, and press both paddle buttons simultaneously to discharge energy to the
patient. For safety reasons, the

SHOCK

button on the defibrillator front panel is disabled when

using standard paddles.

Note: To disarm (cancel the charge), press the Speed Dial. The defibrillator disarms

automatically if shock buttons are not pressed within 60 seconds, or if you change the
energy selection after charging begins. Press

CHARGE

to restart charging.