Therapy

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

4-23

©2006-2013 Physio-Control, Inc.

3 REMOVE TEST PLUG

message appears.

Test plug connected to
QUIK-COMBO therapy cable.

• Disconnect test plug and

connect electrodes to
QUIK-COMBO therapy cable.

Radio frequency interference
during ablation procedure.

• Monitor patient using ECG

cable and therapy cable as
follows:

1 Connect ECG electrodes

to patient and ECG cable.

2 Confirm device displays

patient’s ECG waveform in
Channel 1 in Lead I, II,
or III.

3 Connect Therapy

electrodes to patient and
Therapy cable.

4 Confirm device displays

patient’s ECG waveform in
Channel 2 in Paddles lead.

4

CONNECT CABLE

or

ENERGY NOT DELIVERED

message appears.

Therapy cable became
disconnected and energy was
removed internally.

• Reconnect cable and press

CHARGE

again.

5

ENERGY FAULT

message

appears (selected and
available energy).

Defibrillator is out of
calibration.

• May still transfer energy.
• Contact qualified service

personnel.

6

DISARMING

message

appears.

Discharge button not pressed
within 60 seconds after
charge complete.

• Recharge the defibrillator if

desired.

Energy selected after charge
complete.

• Recharge the defibrillator.

Therapy cable disconnects.

• Reconnect electrode/cable.

7 Patient didn’t jump (no

muscle response) during
defibrillator discharge.

Patient muscle response is
variable and depends on
patient condition. Lack of
visible response to
defibrillation does not
necessarily mean the
discharge did not occur.

• No action needed.

Therapy cable not connected
to therapy electrodes.

• Connect electrodes to

therapy cable.

Table 4-3

Troubleshooting Tips for Defibrillation and Synchronized Cardioversion (Continued)

Observation

Possible Cause

Corrective Action