Physio-Control LIFEPAK 20e User Manual

Page 131

Advertising
background image

7

M

ainta

ining the

Equipment

Maintaining the Equipment

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

7-9

©2006-2013 Physio-Control, Inc.

13 After the tone sounds, indicating full charge, press and hold

SHOCK

while observing the

monitor screen.

14 Confirm the defibrillator discharges on the next sense QRS complex.

15 Confirm the defibrillator returns to asynchronous mode (sense markers are no longer

displayed and sync LED is off).

16 Confirm that the printer annotates the time, date, sync

ON

, sync sense markers prior to

energy delivered, energy selected, no sense markers after

SHOCK

1, and sync

OFF

on the

ECG strip.

17 Connect defibrillator to AC power and power off.

Note: The defibrillator may be configured to remain in synchronous mode after discharge.

Therapy Cable Pacing Check

Equipment needed:

• LIFEPAK 20e defibrillator/monitor

• QUIK-COMBO therapy cable

• Patient ECG cable

• QUIK-COMBO 3- or 12-lead patient simulator

• Fully charged batteries

Procedure:

1 Press

ON

.

2 Connect the QUIK-COMBO therapy cable to the QUIK-COMBO simulator.

3 Turn on the simulator and select

BRADY

.

4 Connect the ECG cable to the defibrillator and the simulator.

5 Select

LEAD II

.

6 Press

PACER

.

7 Confirm that sense markers appear on each QRS complex. If sense markers do not appear,

or appear elsewhere on the ECG, press the selector on waveform Channel 1 and adjust ECG
size on the overlay.

8 Confirm that the Rate overlay appears.

9 Press

CURRENT

and increase the current to 80 mA.

10 Observe the screen for captured complexes. Confirm the pacer LED flashes with each

delivered pacing pulse.

11 Disconnect the QUIK-COMBO therapy cable from the simulator. Confirm that the pacemaker

stops pacing, the

CONNECT ELECTRODES

message appears, and an audible alarm sounds.

12 Reconnect the QUIK-COMBO therapy cable to the simulator. Confirm that the audible alarm

ceases, the

PACING STOPPED

message displays, and that the current is 0 mA.

13 Increase current to 80 mA.

14 Press

CHARGE

. Confirm the pacer LED turns off and the heart rate and available energy

appear on the screen.

WARNING!

Shock hazard.

During defibrillation checks, the discharged energy passes through the cable connectors.
Securely attach cable connectors to the simulator.