Electromagnetic Compatibility Guidance

H-2

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

Table 2

Guidance and Manufacturer’s Declaration - Electromagnetic Immunity

The LIFEPAK 20e defibrillator/monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the LIFEPAK 20e defibrillator/monitor should ensure that the defibrillator/monitor is
used in such an environment.

Immunity Test

IEC 60601

Test Level

Compliance Level

Electromagnetic Environment -

Guidance

Electrostatic
discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.

Electrical fast
transient/burst

IEC 61000-4-4

±2 kV for power supply
lines

±1 kV for input/output
lines

±2 kV for power supply
lines

±1 kV for input/output
lines

Mains power quality should be that of
a typical commercial or hospital
environment.

Surge

IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

±1 kV differential mode

±2 kV common mode

Mains power quality should be that of
a typical commercial or hospital
environment.

Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines

IEC 61000-4-11

<5% U

T

(>95% dip in U

T

)

for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in U

T

)

for 5 s

<5% U

T

(>95% dip in U

T

)

for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in U

T

)

for 5 s

Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
LIFEPAK 20e defibrillator/monitor
requires continued operation during
power mains interruptions, it is
recommended that the LIFEPAK 20e
defibrillator/monitor be powered from
an uninterruptible power supply or a
battery.

Power frequency
(50/60 Hz)
magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.

Note:U

T

is the AC Mains voltage prior to application of the test level.