User test, User test -4, Refer to – Physio-Control LIFEPAK 20e User Manual

Page 126: Figure 7-1

Advertising
background image

Maintaining the Equipment

7-4

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

Figure 7-1

QUIK-COMBO Test Plug

User Test

The LIFEPAK 20e defibrillator/monitor user test performs the same functions as the daily auto
test (refer to the

"Daily Auto Test"

section). The manual user test is recommended if the daily

auto test was not completed, if a test failure was reported, or if REDI-PAK electrodes are
preconnected to the therapy cable as part of defibrillator readiness. This test may also be
performed to meet more frequent defibrillator testing requirements.

To perform the user test, the standard paddles set must be securely seated in the paddle wells or
the QUIK-COMBO therapy cable must be connected to the QUIK-COMBO test plug (see

Figure 7-1

). For REDI-PAK electrodes, disconnect the REDI-PAK electrodes from the

QUIK-COMBO therapy cable and connect the QUIK-COMBO test plug to the therapy cable.

Perform User Test

Procedure:

1 Press

ON

.

Note: To perform the user test, ignore all

REMOVE TEST PLUG

prompts and keep the test

plug connected.

2 Press

OPTIONS

. If the defibrillator is in AED mode, switch to manual mode and then press

OPTIONS

. Refer to

"Switching from AED to Manual Mode," page 4-14

.

3 Select

USER TEST

.

Note: If the message

CONNECT TEST PLUG

appears and the standard paddles set is

securely seated in the paddle wells or the QUIK-COMBO therapy cable is correctly
connected to the QUIK-COMBO test plug, contact a qualified service representative.

4 Select

YES

to initiate user test.

After selecting

YES

, the user test automatically performs the following tasks:

• Performs self-tests.

• Charges to a low energy level (approximately 1-3 joules) and then discharges through a test

plug or hard paddles.

• Tests the pacing circuitry (if noninvasive pacing installed).

• Prints the results.

• Turns itself off.

Note: During the user test, all front panel controls and standard paddles controls are
disabled. Pressing the

ON

button will power off the defibrillator.